R410A
2.9

Pure Blends with Incorrect Component Ratios

The Ultima ID Pro™ has the ability to detect blend ratios that have been altered by
contamination. As stated in section 2.8 Pure Refrigerants, the blends that can be Identified
and Analyzed are R22, R32, R134a, Hydrocarbons (HC), R404A, R407C and R410A. If
one of the multi-component refrigerant blends has incorrect component ratios, it will display
as shown below in (Figure 17). To see the component concentrations, press the 'MORE'
button (Figure 18). Refer to section 2.10 Print the Test Results & Channel Data to print.
COMPONENTS OF
R410A

2.10 Measuring Air

The Ultima ID Pro™ offers the ability to measure the presence of Air independently during
every test. The built in oxygen sensor will display the percentage of Air in all Identified or
Identified and Analyzed refrigerants if there is Air in a system or cylinder. The image below in
(Figure 19) portrays how the LCD will display Air if detected. Pressing 'MORE' will display
the purity of the refrigerants which were detected (Figure 19a).
NOTE:
Air is measured independently from refrigerant! You can have 100% pure
refrigerant with a percentage of Air in a system or tank. The presence of Air will display on
the initial results screen and on every printed test.
R22
Manual Part Number:
5-06-7000-70-2
0.0% AIR
100%
MORE
Figure 15
0.0% AIR
MORE
Figure 17
9.6% AIR
PRINT
MORE
Figure 19
COMPONENTS OF
R410A
R125
R32
BACK
COMPONENTS OF
R410A
R125
R32
BACK
R22
BACK
Manual File:
MN-A-0289 Rev. E
RESULTS -
-
100%
50.0%
50.0%
PRINT
DONE
Figure 16
RESULTS -
-
100.0%
60.0%
40.0%
PRINT
DONE
Figure 18
- RESULTS –
100.0%
PRINT
DONE
Figure 19a
Page 2-20