Instructions For Use - Ambu Oval Plus Silicone Resuscitator Instrucciones De Uso

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 54
Neonate
Resuscitator volume
approx. 260 ml
Delivered volume one hand**
150 ml
Delivered volume two hands**
-
Dimensions without
approx.
Reservoir (length x diameter)
198 x 75 mm
Weight without
approx. 128 g
Reservoir and Mask
Pressure-limiting valve*
4.0 kPa (40 cm H
≤ 5 ml + 10 % of
Dead space
the delivered volume
max 0,5 kPa
Inspiratory resistance
(5,0 cm H
at 5 l/min
max 0,5 kPa
Expiratory resistance
(5,0 cm H
at 5 l/min
approx. 1500 ml (bag)
Reservoir volume
approx. 100 ml (tube)
Patient connector
Expiration connector
(for PEEP valve attachment)
Manometer Port connector
Bag inlet connector
-
Forward and backward leakage
Operation temperature
Tested at - 40 °C (-40 °F) and + 60 °C (+140 °F)
Storage temperature
O
inlet connector
2
For long term storage the resuscitator should be kept in closed
Long term storage
packaging in a cool place away from sunlight.
Paediatric
Adult
approx. 660 ml
approx. 1546 ml
450 ml
600 ml
-
1100 ml
approx.
approx.
240 x 100 mm
269 x 130 mm
approx. 205 g
approx. 338 g
O)
4.0 kPa (40 cm H
O)
4.0 kPa (40 cm H
2
2
≤ 5 ml + 10 % of
≤ 5 ml + 10 % of
the delivered volume
the delivered volume
max 0,5 kPa
max 0,5 kPa
O)
(5,0 cm H
O)
(5,0 cm H
2
2
2
at 50 l/min
at 50 l/min
max 0,5 kPa
max 0,5 kPa
O)
(5,0 cm H
O)
(5,0 cm H
2
2
2
at 50 l/min
at 50 l/min
approx. 1500 ml (bag) approx. 1500 ml (bag)
Outside 22 mm male (ISO 5356-1)
Inside 15 mm female (ISO 5356-1)
30 mm male (ISO 5356-1)
Ø 4,2 +/- 0,1 mm
Inside 32mm female
Inside 32mm female
(ISO10651-4)
(ISO10651-4)
Not measurable
-18 °C to +50 °C (-0.4 °F to +122 °F)
according to EN ISO10651-4
according to EN 13544-2
* Higher delivery pressure can be obtained by overriding the pressure-limiting valve.
The Ambu Oval Plus Adult is also available without pressure-limiting valve.
** Tested according to EN ISO 10651-4.
4. Principle in operation 
The illustration (1) shows the ventilation gas flow mixtures into the bag and to and
from the patient during manual operation of the resuscitator. (a) Adult and paedi-
atric resuscitator, (b) infant resuscitator with closed reservoir, (c) infant resuscitator
O)
2
with open reservoir.
The gas flow is similar when the patient is breathing spontaneously through the
device. The O
reservoir assembly is fitted with two valves, one allowing ambient
2
O)
air to be drawn in when the reservoir is empty and one spilling out surplus
oxygen when the reservoir bag is full.
1.1 Excess oxygen, 1.2 Air, 1.3 Oxygen inlet, 1.4 Patient, 1.5 Expiration,
O)
1.6 Manometer port, 1.7 Pressure limiting valve,

5. Instructions for use

Preparation
- If the resuscitator is packed in a compressed state, unfold by pulling on
the patient valve and the inlet valve.
- Fit the facemask and place all items in the plastic bag supplied with
the resuscitator.
- The integrity of the kits issued for storage ready for use should be inspected
at the interval established in the local protocol.
- Before use on the patient make a brief functional test as described in section 9.
- If connecting external devices to the resuscitator, make sure to test for functionality
and consult the instructions for use accompanying the external device.