Oxygen enrichment of the inspiratory air in case of
spontaneous respiration
• With oxygen reservoir bag or without an oxygen reservoir:
Attach a mask to the resuscitator, set the oxygen flow to between 2 and 4 l/min,
and place the mask over the patient's mouth and nose, creating an airtight seal.
If the spontaneous respiration is sufficient, an oxygen concentration close to 100
% will be obtained.
• With an oxygen reservoir tube:
The end of the resuscitator with the oxygen reservoir that normally faces away
from the patient can also be used for oxygen enrichment of the inspiratory air in
case of spontaneous respiration. Set the oxygen flow should be set to between
2 and 4 l/min, and the free end of the oxygen reservoir should be held as tightly
as possible against the patient's head.
The oxygen concentration will be very close to 100 % at the end of the tube.
Depending on the distance between the oxygen reservoir and the patient's head,
there will be significantly more or less inhalation of atmospheric air, and thus
lower oxygen concentrations should be expected.
7. Parts/Materials
Bag
Patient valve housing
O-Ring (only patient valves with swivel)
Patient connector
Expiratory connector
Outlet cap (option)
Valve discs
Reservoir bag
Silicone rubber
Polysulphone
Silicone rubber
Polysulphone
Polypropylene, reinforced
Polysulphone
Silicone rubber
Polyethersulfone/Polyurethane
Flange nut
Pressure limiting valve
Override cap
Bag support connector
Inlet valve cover
Inlet valve housing
Cap for manometer port
Inlet valve cover (Neonate)
Inlet valve housing (Neonate)
Window with snap faster
Reservoir housing
Adapter for reservoir bag
Reservoir tube
8. Cleaning-disinfection-sterilisation
Clean – disinfect – sterilize the complete resuscitator thoroughly after each
patient use or if used in a patient/environment with infectious diseases.
8.1 Disassembly
The resuscitator must always be disassembled for cleaning as shown.
Adult (6.1), Peadiatric (6.2), Neonate (6.3)
CAUTION
Do not try to disconnect the patient connector from the patient valve
(permanent snap fit).
Do not disassemble parts further than shown.
The bag connector should not be detached from the bag during
the cleaning and sterilisation procedure.
Polypropylene, reinforced
Polysulphone/stainless steel
Silicone rubber
Polysulphone
Polyoxymethylene
Polyoxymethylene
Polypropylene/EPDM rubber
Polysulphone
Polysulphone
Polysulphone
Polyoxymethylene/Polypropylene
Silicone rubber
Polyethylene