8.
[Samp] key (COUNTING MODE):
a) Turn on the scale and wait a few seconds.
b) Put on the plate a fix number of pieces, for instance 10 pieces (the more,
the better).
c) Select with the buttons [0,1,....9] the number of pieces you put on the
plate
d) Press [sample]. On the display [NUMBER], the number of pieces is shown,
and on the display [UNIT WEIGHT], the weight of one piece will be
calculated and shown (in grams). On the display [WEIGHT], the weight of all
the samples will be shown.
e) The counting mode is ready for use.
9.
Backlight On/Off: Press [Zero] key for 3-4 seconds, the backlight will be switched
on or off.
10. Adjust the decimal point number for the unit weight: Enter the unit weight form
the keyboard, when the unit weight is zero, press number key "0" to adjust the
decimal point number for the unit weight.

ALARM INDICATION

1. The weight window displays "-----" with continuously alarm sound when
EN
theweight over 100%FS+9d, if the ADC overflows, the weight window displays
"—Adc—"with continuously alarm sound. For these instance, the weight
should be taken away.
2. The number window displays "—OF—" when the number or accumulated
number over 99999. The scale will not response if the accumulated times over
100, but the original accumulated data will be retained.
3. If the voltage of the accumulator is low, the weight window displays "—Lb—"
while the weight is zero (The weight display will recover to normal when it is
loaded ). Under this condition, you can use the scale in a short time, but you
should plug in the AC plug as soon as possible to recharge the accumulator.
4. The weight window displays "HHHH" or "LLLL" when the zero weight is
higher or lower than the permitted range.
5. If the weight window displays "-SYS-", it means the system parameter is wrong.,
you should send it for maintain immediately.
6. If the weight window displays "Err-2" after self-testing, it means unstable
weighing, please see III/1/(2),(3).
8
Rev. 01.07.2020

WEIGHT CALIBRATION

Turn on the scale and wait a few seconds. Press and hold ACCU button as long as
UNIT WEIGHT window will show "LoAd" and WEIGHT window will show 30.000.
Put on the plate 30 kg. When the scale will be stable, press ACCU button. Calibration
is finished, scale goes to normal operation mode.
REGULAR TROUBLESHOOTING
1. Unstable weighing
Low voltage for accumulator: Switch off the scale immediately and recharge it
more than 12 hours.
Low intensity or strong shake for platform: Strengthen the platform or avoidshake.
Influenced by the wind: Set a glass cover to reduce the influence of the wind.
2. No display when switch on the scale
Accumulator is unusable: Change the accumulator if it remains invalid after
recharging.
Broken with fuse or switch: Change the fuse or switch.
Others: Send it for maintain.

TRANSPORTATION AND STORAGE

Avoid shock and dampness during transportation.
Avoid strong sunshine and rain during storage.
ATTENTION
1.
Please read this instruction manual carefully before you use the scale.
2.
If it has not be used for long time, please recharge it once 2 months at least,
andkeep the power switch on "off" status to avoid discharge of accumulator.
3.
LCD of the scale must to be operated under rich light, you are advised to select
product with backlight under low light condition.
4.
Avoid collision when you put the weighting object on the scale.
5.
It is strictly prohibited to use strong chemical solvent to clean the surface.
6.
Please clean the scale with dry or wet dishcloth. Washing it with water is
strictlyprohibited.
7.
Please don't use the same power supply circuit together with the big power
equipment, because the running of the big power equipment will influence the
weighing.
CAUTION
If the weight window displays "—Lb—" when the weight is zero, it means the
voltage of the accumulator is low. Under this condition, you can use the scale in a
short time, but you should plug in the AC plug as soon as possible to recharge the
accumulator, please recharge the accumulator for more than 12 hours.
The accumulator is not included in free maintain warrant.
Rev. 01.07.2020
9