Single-use bonnets
Material
Polyamide
Use
Single use
Available sizes:
XXS - XXXL
Combinable nCPAP generators: Medijet 1000, Medijet 1010, Medijet 1020 and Miniflow 4000
3.
Set-up and application
Necessary parts:
To administer effective nCPAP therapy with Miniflow, the following equipment is necessary:
•
A ventilator or a bubble CPAP system
•
An active respiratory gas humidifier
•
a suitable circuit set including inspiratory and expiratory tubes and a chamber for the humidifier. The connectors at the
end of the inspiratory and the expiratory tubes have to be 10 female to fit Miniflow.
•
A measuring tape to measure the necessary bonnet and prong size
•
a bonnet that fits the baby
•
a mask or prong of the correct size
•
optional for ventilators which require external pressure measurement:
a pressure measuring T-piece already included inside medin's circuits or ordered separately (REF 4010)
Warning: packaging
If the packaging of any part or the part itself is damaged, this part cannot be used and must be discarded and disposed of.
Also check that the holes of the generator and of the prongs /masks are not blocked. In the event of blockage, the products in
question must not be used.
Structure and set-up of the system (disposable bonnets):
(1) Connect the circuit set to the ventilator and humidifier. If
necessary, use the extra connectors. Insert the temperature
sensors completely, until the tip reaches the gas flow.
(2) Connect Miniflow to the inspiratory and expiratory tubes
If your ventilator requires external pressure measurement,
insert the T piece for pressure measurement in the expiration
part, where you can insert the pressure line and connect it
to the ventilator.
(3) Close the prong adapter of the Miniflow with your finger
and adjust the nCPAP parameters on the ventilator, e.g.
the pressure. If this is not possible, check the connections
to see if one is leaking. If all connections are correct and
Miniflow is the reason for the failure, discard this Miniflow
and dispose of it.
Important instruction: defective Miniflow
If Miniflow is not working correctly, it may not be used.
7
Masks
Silicone
Single use
XS - XL
User Manual Miniflow
Prongs
Silicone
Single use
XS - XL