4. Move the movable parts back and forth (at least 5 times).
5. Brush the inside of the shaft tube with a round brush (at
least 5 times).
6. Rinse the surfaces with sterile, deionized water (at least
1 minute).
7. Completely fill a disposable syringe with sterile, deionized
water and rinse the lumina (at least 5 times).
8. Repeat the preceding cleaning steps if the product or used
rinse water still contains visible contamination.
Disinfection
1. Make sure the instrument jaws are open.
2. Immerse the product completely in a disinfectant bath, en-
suring that the product does not come into contact with oth-
er components in the bath.
3. Move the movable parts back and forth (at least 5 times).
4. Completely fill a disposable syringe with disinfecting solution
and rinse the lumina (at least 5 times).
5. Rinse the surfaces with sterile, deionized water (at least
1 minute).
6. Completely fill a disposable syringe with sterile, deionized
water and rinse the lumina (at least 5 times).
7. Disconnect the irrigation tube from the shaft tube.
8. Dry the product, including the lumens, until there are no
more fluid residues visible (e.g. with filtered compressed air).
7.7
Cleaning and disinfection by machine
The washer/disinfector must have fundamentally certified ef-
fectiveness (e.g. CE marking according to DIN EN ISO 15883).
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 12.3 (handles: max. 11!).