WIKA TR10-0 Manual De Instrucciones página 22

Ocultar thumbs Ver también para TR10-0:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

11. Specifications
11.2 Model TC10-0
Measuring element
EN
Type of measuring
element
Measuring current
Validity limits of the class accuracy per EN 60584-1
Type K
Type J
Type E
Type N
Type T
Validity limits of the class accuracy in accordance with ASTM-E230
Type K
Type J
Type E
Type N
Type T
For further specifications see WIKA data sheet TE 66.01 and the order documentation.
22
Thermocouple per IEC 60584-1 or ASTM E230
Types K, J, E, N, T
Ungrounded welded (standard)
Welded to the bottom (grounded)
Class 2
-40 ... +1.200 °C [-40 ... +2,192 °F]
Class 1
-40 ... +1.000 °C [-40 ... +1,832 °F]
Class 2
-40 ... +750 °C [-40 ... +1,382 °F]
Class 1
-40 ... +750 °C [-40 ... +1,382 °F]
Class 2
-40 ... +900 °C [-40 ... +1,652 °F]
Class 1
-40 ... +800 °C [-40 ... +1,472 °F]
Class 2
-40 ... +1,200 °C [-40 ... +2,192 °F]
Class 1
-40 ... +1,000 °C [-40 ... +1,832 °F]
Class 2
-40 ... +350 °C [-40 ... +662 °F]
Class 1
-40 ... +350 °C [-40 ... +662 °F]
Standard
0 ... 1,260 °C [32 ... 2,300 °F]
Special
0 ... 1,260 °C [32 ... 2,300 °F]
Standard
0 ... 760 °C [32 ... 1,400 °F]
Special
0 ... 760 °C [32 ... 1,400 °F]
Standard
0 ... 870 °C [32 ... 1,598 °F]
Special
0 ... 870 °C [32 ... 1,598 °F]
Standard
0 ... 1,260 °C [32 ... 2,300 °F]
Special
0 ... 1,260 °C [32 ... 2,300 °F]
Standard
0 ... 370 °C [32 ... 698 °F]
Special
0 ... 370 °C [32 ... 698 °F]
WIKA operating instructions models TR10-0, TC10-0 (FM, CSA)
Tabla de contenido
loading

Este manual también es adecuado para:

Tc10-0

Tabla de contenido