Remove the tip ejector. Unscrew the nut, remove
the shaft, piston assembly, seal and O-ring. Rinse
the removed parts with distilled water and dry thor-
oughly. Lightly lubricate the inside surfaces of the
seal and the O-ring with the lubricant. Reassemble
the pipette in the reverse order.
Note: The shaft, shaft nut, ejector, piston assembly,
seal and O-ring may be autoclaved at a temperature
of 121°C for 20 minutes at pressure 1 bar.
The shaft of the 5000 and 10000 models should be
autoclaved without the filter.
The handle and the outside of the pipette may be
cleaned using a cloth damped in isopropanol.
If the pipette malfunction persists after carring out the
above steps, send the pipette to your HTL service repre-
sentative.
Before returning the pipette, make sure that it is free from
any chemical, radioactive or microbiological contamina-
tion that might pose any danger during transportation
and repairing. Clean the pipette as thoroughly as possi-
ble.
10 - PIPETTE KIT
The pipettes are delivered in the kits including:
• Pipette
• Instruction manual
• Calibration key
• Pipette stand
• Ejector regulation spacers
(for pipette models LM2 - LM1000)
• Ejector cap (for pipette models LM2, LM10, LM5000,
LM10000)
• Tips
• Identification labels
• Filters (for pipette models LM5000, LM10000)
• Lubricant
The stand assembly diagram is shown in Fig. 5.
13
11 - SPARE PARTS
All the spare parts indicated in Fig. 3, 4, 6 that is:
A: Pushbutton
A1: Pushbutton
B: Calibration knob
C: Shaft
D: Ejector
F: Shaft nut
G: Piston assembly
H: Spacer
I: O-ring
J: Seal
K: Calibration key
L: Filter
M: Ejector cap
N: Ejector pushbutton
can be ordered from your HTL representative (type of
pipette and name of the part for this pipette should be
specified).
Warning: The replacement of the piston requires
conducting of calibration procedure according to
section 8.