Pipette Kit; Spare Parts - HTL Discovery Comfort DV2 Manual De Instrucciones

Ocultar thumbs Ver también para Discovery Comfort DV2:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 29
It is recommended:
- to sterilize the pipettes in autoclave with an initial vacu-
um and drying cycle,
- prior to sterilization unscrew the shaft nut slightly in the
DV/D2-DV/D1000 pipettes, and unscrew the shaft
slightly in the DV/D5000 and DV/D10000. After autoclav-
ing these parts should be screwed tight again.
- to set the locking ring in lower (unlocked) position prior
to sterilization
The precision of the results should not alter if the pipet-
ting process and autoclaving are carried out as descri -
bed in this manual. Because a slight change in the accu-
racy of the dosage may occur, it is recommended to:
- check the calibration of the pipette after the initial first,
third and fifth autoclaving cycles and then after every 10
autoclaving cycles.
The pipettes are UV resistant which was confirmed by
our tests. The recommended distance from the radiation
source to exposed element should be not less than
50 cm. Prolonged and very intense UV exposure can
cause de-coloration of pipette parts, without affecting
its performance.

11 - PIPETTE KIT

The pipettes are delivered in the kits including:
• Pipette
• Instruction manual
• Calibration key
• Pipette stand
• Ejector cap (for pipette models DV/D2, DV/D10,
DV/D5000, DV/D10000)
• Identification labels
• Filters (for pipette models DV/D5000, DV/D10000)
• Lubricant
• Color identification rings
(for pipette models DV/D2 - DV/D1000)
The stand assembly diagram is shown in Fig. 8
15

12 - SPARE PARTS

The spare parts indicated in Fig. 3, 4, 5 and 7 that is:
A: Pipetting pushbutton
B: Adjustment knob
C: Shaft
D: Ejector
F: Shaft nut
G: Piston assembly
I: O-ring
J: Seal
K: Calibration key
L: Filter
M: Ejector cap
N: Ejector pushbutton
P: Identification ring
S: Locking ring
can be ordered from your HTL representative (type of
pipette and name of the part for this pipette should be
specified).
Warning: The replacement of the piston assembly must
be followed with calibration procedure (see section 8)
All rights reserved. Product described in this manual is subject to availa-
bility and technical modification. Errors excepted.
PZ HTL S.A. reserves the right to improve, enhance or otherwise modify
its products without prior notification.
© 2015 PZ HTL S.A.
ENGLISH
A1: Pushbutton
A2: Knob
16
Tabla de contenido