nominal volume of the pipette, exceeds the per-
missible value given in the table in section 1,
you should recalibrate the pipette according to
the rules presented in section 5.
Increased ejector pushbutton turning resis -
tan ce.
• The handle (5) inner surface is dirty.
Remove the ejector (16), pressing the ejector
push-button (2) unscrew the nut (14), pull out
the ejector spring (7). Slide the body (6) out
from the handle (5).
Clean the handle inner surface and the body
outer surface with dry cotton wool. Grease the
ejector spring, handle inner surface and the
body outer surface which enters the handle,
with lubricant.
Re-assemble the pipette in the reverse order.
Attention: Use only lubricant that has been
included with each pipette.
If the pipette malfunction persists, deliver it to
your HTL service representative.
Before returning the pipette make sure that it is
free from any chemicals, radioactive or micro -
bio logical contamination that might pose any
danger during transportation and servicing.
If possible, clean the pipette to perform the ste -
ry lisation.
11
8 - CLEANING AND STERILISATION
Cleaning:
External surfaces of the pipetting pushbutton (1)
- Fig. 6, the ejector pushbutton (2), the handgrip
(5) and the body (6) can be cleaned with tissue
dipped in isopropyl alcohol. Remaining parts
removed from the pipette during disassembly (7
to 17) can be washed with distilled water or iso-
propyl alcohol.
Sterilisation:
The pipette can be subjected to sterilisation, as
a whole, in the autoclave at temperature 121°C
for 20 minutes. After performing the sterilisation
the pipette should be dried and cooled down to
room temperature.
It is recommended:
- to sterilise the pipettes in an autoclave with
initial vacuum and drying cycle,
- to test pipette calibration every 10 sterilisation
cycles.
The pipettes are UV resistant which was con-
firmed by our tests. The recommended distance
from the radiation source to exposed element
should be not less than 50 cm. Prolonged and
very intense UV exposure can cause de-col-
oration of pipette parts, without affecting its per-
formance.
9 - PIPETTE KIT
Pipettes are delivered in the kits including:
• Pipette,
• Instruction manual,
• Calibration key,
• Pipette stand,
• Lubricant.
The way of fixing the stand is shown in Fig. 5.
ENGLISH
12