Installation
WARNINGS:
• Materials of construction are chosen based on their chemical compatibility (or
inertness) for general purposes. For exposure to specific environments, check with
chemical compatibility charts before installing.
• Explosion hazard. Substitution of components may impair suitability for Class I,
Division 2 applications.
• This product can only function properly and safely if it is correctly transported,
stored, installed, set up, operated, and maintained.
• The user is responsible for the selection of bolting and gasket materials which will
fall within the limits of the flange and its intended use, and which are suitable for
the service conditions.
Installation shall only be performed by qualified personnel and in accordance with
Note:
local governing regulations.
Mounting Location
Recommendations:
• Ambient temperature within -40 to +60 °C (-40 to +140 °F), -20 °C (-4 °F) if metal mounting.
• Easy access for viewing the display and programming via two push buttons.
• An environment suitable to the housing rating and materials of construction.
• Keep the sound path perpendicular to the material surface.
Precautions:
• Avoid proximity to high voltage or current wiring, high voltage or current contacts, and to
variable frequency motor speed controllers.
• Avoid interference to the sound path from obstructions or from the fill path.
The sound path should be:
•
perpendicular to the
monitored surface
•
clear of rough walls,
seams, rungs, or
other obstructions
•
clear of the fill path
Mounting Instructions
Ideally, mount Pointek ULS200 so that the face of the transducer is at least
Note:
250 mm (9.84") above the highest anticipated level.
Pointek ULS200 is available in three thread types: 2" NPT, 2" BSPT, or 2" G.
Before inserting Pointek ULS200 into its mounting connection, ensure that the threads are of
the same type to avoid damaging them. Simply screw Pointek ULS200 into the process
connection, and hand tighten.
A5E32268616
Pointek ULS200 – QUICK START MANUAL
pipe
rungs
seams
fill
Page EN-5