Problem
The ResMed S7 Elite unit does
not stop when you remove
your mask.
SmartStart is enabled but the
flow generator does not stop
automatically when you
remove your mask.
Display error message:
Check tube!!
Key if done
Display error message:
SYSTEM ERROR
Call service!
Display error message: High
leak in last session.
30
Possible Cause
SmartStart/Stop is disabled.
Incompatible humidifier or mask
system being used.
The air tubing is loose.
Component failure.
You have experienced excessively
high leak levels during the night.
Solution
Enable SmartStart/Stop.
Use only equipment as
recommended and supplied by
ResMed.
Check that the air tubing is
connected securely to your mask
and the air outlet on the front of
the ResMed S7 Elite.
When you have checked the air
tubing, press the Front key to
restart the machine.
If this does not clear the message,
turn the power switch off and then
turn it back on to restart the
machine.
Turn your ResMed S7 Elite off and
then turn it back on. If the error
message is still displayed, return
the device for servicing.
Check that your air tubing is
connected properly and that your
mask does not leak excessively.
Use the mask-fitting feature to
help you to fit your mask properly.
If this message appears again,
contact your clinician.