Operating the SimplyGo Mini Device

Power Options

Warning: Do not use the device without the battery installed and operating�
To run the device on battery power, make sure the battery is properly
inserted and sufficiently charged� (The battery should be fully
charged the first time you use it�)
To run the device using AC power, first make sure you have a
battery installed� Then, follow these steps:
1�
Connect the output cable of the AC Power Supply to the
SimplyGo Mini power input connector, aligning the green dots
on the power cord and device as shown below�
2�
Connect the AC Power Cord's connector into the AC power
supply�
3�
Connect the AC Power Cord's plug into an AC outlet installed to
meet national or regional electrical code�
To run the device using DC power from an automobile or other
motorized vehicle, make sure the battery is installed� Start the vehicle
and connect one end of the DC Power Cord to the device's power
inlet and the other end into the vehicle's DC electrical outlet�

SimplyGo Mini Power Cord Connection

When connecting the SimplyGo Mini
device to either an AC or DC power supply:
1�
Align the green dot indicators
on the power cord�
2�
Insert the cord connection
into the device as shown in
the illustration�
Note: DC power provided by the DC
Power Supply will also charge the battery when the device is turned off
or is running on setting 1, 2, or 3� It may not charge the battery when
the device is running on setting 4 or 5� Charging status can be verified
with the battery icon on the Home Screen�