SICK SIDOR E Instrucciones De Operación página 16

Tabla de contenido
2
ENGLISH
16
A D D E N D U M T O O P E R A T I N G I N S T R U C T I O N S | SIDOR E
Press key "2" to go to the "end value" menu. You can set the desired measuring range
9
there.
The allowed range for the measuring range is specified by "min. value" and "max.
value".
The active setting is shown as "status", 96.53 in the example below.
settings 621122
end value
CO
unit: mg/m3
min. value:
10.00
max. value: 100.00
Status: 96.53
Input:
Save: ENTER
Back: ESCAPE
No messages
CO
12 mg/m3
10
Enter the desired measuring range, "80" in the example below.
settings 621122
end value
CO
unit: mg/m3
min. value:
10.00
max. value: 100.00
Status: 96.53
Input:
80
Save: ENTER
Back: ESCAPE
No messages
CO
12 mg/m3
Press "ENTER" to save.
11
Press "ESC" several times to return to the main menu.
12
8019968/V1-0/2016-06 | SICK