This alarm converts a therapy pause into an alarm if the duration of the pause is too long.
Alarm generation:
When paused for five minutes, the product will display the Therapy Paused Time Out Alarm. After you resume the current
therapy, the Therapy Paused Time Out Alarm is deactivated.
N N o o t t e e - - The audio paused timeout is 10 minutes for the therapy paused time out alarm.
R R e e m m o o v v e e f f r r o o m m u u s s e e m m o o d d e e
The Remove From Use (RFU) is a safety mode to limit operations. A fault condition prevents the product from normal
functions and requires service. The controller will stop the Active Therapy and communicate to the operator that the
controller is going into RFU mode.
C C A A U U T T I I O O N N - - Always remove the product from use before servicing any components. Contact qualified service personnel for
service.
Depending on the remove from use (RFU) condition, text may or may not be displayed. For example, if there is a power
loss.
Water temperature probes are out of the allowed range
Program and data checksum error
High thermal cutout test failed
Backup power product replacement required
Low or over safety temperature
Pump over current
Compressor power fault
Heater power fault
Refrigerant control valve fault
Main DC power lost
CAN heart beat lost
Dual safety temperature sensors do not match readings
Dual safety temperature sensors are out of the allowed range
Hardware watchdog heartbeat failure
N N o o t t e e - - The audio paused timeout is 10 minutes for the remove from use alarm.
8001-109-005 Rev C.2
57
EN