S S w w i i t t c c h h i i n n g g m m o o d d e e s s
Tap Edit and select a different therapy mode.
P P a a u u s s i i n n g g a a n n d d r r e e s s u u m m i i n n g g t t h h e e r r a a p p y y
To pause therapy, press and hold the Pause Therapy button for two seconds.
To resume therapy, press and hold the Pause Therapy button for two seconds.
D D i i s s p p l l a a y y i i n n g g t t h h e e d d a a t t a a s s t t o o r r a a g g e e
The system gathers data at 60 second intervals and is limited to 90 minutes of storage. The graphical display defaults to
show data for all four variables in the Manual and Automatic modes.
To display the patient data graph:
1. Tap the Graph icon.
Primary patient temperature reading from attached probe (A) (Figure 19)
Intermediate target temperature (B)
Water temperature (C)
Power level (D)
2. To view data values, hide values, or data lines, tap an icon until the data you desire appears for the selected icon.
3. Tap next to see the current values for each variable.
4. To exit, tap the Graph icon or the Exit button.
N N o o t t e e
• In monitor mode, only the patient temperature data is displayed (A).
• The graph icon is only available when a therapy is active.
• The patient data remains until the product sleeps or you powered off the product.
• If you lose power for more than 10 minutes, the data will not be retrievable.
8001-109-005 Rev C.2
F F i i g g u u r r e e 1 1 9 9 – – G G r r a a p p h h i i c c a a l l d d i i s s p p l l a a y y
A
35.4°C
B
C
8.0°C
D
0
31