21. Pull up on the controller drain plug (Figure 32) to open the drain.
A
F F i i g g u u r r e e 3 3 2 2 – – D D r r a a i i n n p p l l u u g g
22. Make sure that all 3 hoses remain raised above the connection ports for draining.
23. Allow the controller and hoses to drain for a minimum of two minutes.
24. Push down on the controller drain plug to close the drain.
25. When the controller and hoses are drained, continue to Rinsing the internal water circuit and hoses (page 44).
R R i i n n s s i i n n g g t t h h e e i i n n t t e e r r n n a a l l w w a a t t e e r r c c i i r r c c u u i i t t a a n n d d h h o o s s e e s s
1. Using appropriate measuring equipment, fill the empty reservoir with 1 gallons (3.8 L) of sterile distilled water or distilled
water that has been passed through a filter less than or equal to 0.22 microns.
2. Place the reservoir into the controller.
3. Connect the bottom hose end to the hydraulic connector in the lid of the reservoir.
F F i i g g u u r r e e 3 3 3 3 – – B B o o t t t t o o m m h h o o s s e e e e n n d d i i n n t t h h e e l l i i d d o o f f t t h h e e r r e e s s e e r r v v o o i i r r
4. Plug the power cord into a wall outlet.
5. Press and hold the
Stand-by button.
6. Tap the Manual mode icon.
7. Tap Confirm.
8. Select the water target temperature of 25.0° C (77.0° F).
EN
44