9. Discard the disposable thermal transfer devices based on your local waste management protocol.
a. Discard the disposable thermal transfer devices based on your local waste management protocol, if applicable.
D D r r a a i i n n i i n n g g w w a a t t e e r r f f r r o o m m t t h h e e r r e e s s e e r r v v o o i i r r
To drain the water from the reservoir:
1. See Removing and replacing the reservoir (page 26).
2. Dispose of the water per hospital protocol.
3. Replace the reservoir.
N N o o t t e e - - Make sure that the reservoir is dry before you store the product.
D D r r a a i i n n i i n n g g w w a a t t e e r r f f r r o o m m t t h h e e c c o o n n t t r r o o l l l l e e r r a a n n d d h h o o s s e e s s
Make sure that the controller and all components are dry before you store the product. Make sure to drain the hoses before
you store them.
1. Place the controller over a floor drain.
2. Remove the reservoir and pull up on the controller drain plug (A) to open the drain (Figure 21).
3. Connect a hose to each port.
a. If you have Colder style connector hoses, attach the service tool adapter hose (8001-999-017).
b. If you have C C l l i i k k - - T T i i t t e e hoses, make sure that the connectors and clamps are open (Figure 22).
8001-109-005 Rev C.2
A
F F i i g g u u r r e e 2 2 1 1 – – D D r r a a i i n n p p l l u u g g
33
EN