Stryker 8001 Altrix Manual Del Usuario página 47

Ocultar thumbs Ver también para 8001 Altrix:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 271
F F i i g g u u r r e e 3 3 0 0 – – D D i i s s c c o o n n n n e e c c t t e e d d h h o o s s e e
6. Connect the bottom hose end to the hydraulic connector in the lid of the reservoir (Figure 31).
F F i i g g u u r r e e 3 3 1 1 – – B B o o t t t t o o m m h h o o s s e e e e n n d d i i n n t t h h e e l l i i d d o o f f t t h h e e r r e e s s e e r r v v o o i i r r
7. Plug the power cord into a wall outlet.
8. Press and hold the
Stand-by button.
9. Tap the Manual mode icon.
10. Tap Confirm.
11. Set the water target temperature to 25.0° C (77.0° F).
12. Tap Confirm.
13. Allow the controller to run for 2 minutes to purge air from the system.
1
14. Run the controller for 20 minutes.
15. After 22 minutes, turn the controller off by pressing and holding the Stand-by button for two seconds.
16. Unplug the power cord from the wall outlet.
17. Place the controller over a floor drain.
18. Remove the reservoir. Pull forward at an angle, and lift out the reservoir.
19. Remove the bottom hose end from the hydraulic connector adapter in the reservoir lid by pushing down on the collar.
20. Empty water from the reservoir, dispose of the water per hospital protocol.
N N o o t t e e - - Do not rinse the reservoir.
8001-109-005 Rev C.2
43
EN
Tabla de contenido
loading

Tabla de contenido