D D i i s s i i n n f f e e c c t t t t h h e e i i n n t t e e r r n n a a l l w w a a t t e e r r c c i i r r c c u u i i t t a a n n d d h h o o s s e e s s e e v v e e r r y y 1 1 4 4 d d a a y y s s
Use the B B r r u u C C l l e e a a n n T T b b C C disinfectant tablets manufactured by B B r r u u l l i i n n (EPA registration number 71847–2–106) or equivalent,
before first use, at least every 14 days, and before storage. B B r r u u C C l l e e a a n n T T b b C C has been validated for internal water circuit
disinfection. Make sure that you follow the disinfectant manufacturer's guidelines to avoid the risk of injury. Failure to follow
the disinfectant's instructions may void your warranty.
C C A A U U T T I I O O N N
• Always use sterile distilled water or distilled water that has been passed through a filter less than or equal to 0.22
microns with this product.
• Do not disinfect the internal water system with a thermal transfer device attached as this may cause a leak.
• Do not use bleach or any other cleaning or disinfectant agents for internal circuits. This could result in damage to the
product. Only use approved disinfectant tablets.
• Always drain the product before disinfecting the internal water circuit. Failure to drain the product may reduce the
effectiveness of the disinfection process.
N N o o t t e e - - Disinfection of the A A l l t t r r i i x x internal water system was validated using M. mucogenicum .
T T o o o o l l s s R R e e q q u u i i r r e e d d : :
2 gallons (7.6 L) of sterile distilled water or distilled water that has been passed through a filter less than or equal to 0.22
microns
Personal protection equipment (PPE) as recommended by the disinfectant manufacturer's instructions
Soft, lint free cloth (2 or more)
2 B B r r u u C C l l e e a a n n T T b b C C tablets (active ingredient is sodium dichloroisocynaturate (NaDCC); 1874 ppm disinfectant solution
when dissolved in one gallon of water) or equivalent
Service tool adapter hose (8001-999-017) for C C o o l l d d e e r r style connector hoses
Floor drain
See Product illustration (page 10) for clarification of product component names and locations.
D D r r a a i i n n i i n n g g t t h h e e i i n n t t e e r r n n a a l l w w a a t t e e r r c c i i r r c c u u i i t t a a n n d d h h o o s s e e s s f f o o r r d d i i s s i i n n f f e e c c t t i i o o n n
1. Unplug the power cord from the wall outlet.
2. Place the controller over a floor drain.
N N o o t t e e - - For best results, the floor drain should be within reach of a wall outlet to power on the controller.
3. To drain the controller, pull up on the controller drain plug (A) to open the drain (Figure 25). Leave the drain open.
EN