Cleaning And Disinfecting Peripheral Devices - SonoSite SiteStand Manual Do Usuário

Sistema do acoplamento móvel
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 76
To clean and disinfect the SiteStand:
The exterior surface of the SiteStand can be cleaned and disinfected using a recommended cleaner
or disinfectant.
Remove the system and all contents from the SiteStand.
1
Use a soft cloth lightly dampened in a mild soap or detergent cleaning solution to clean the
2
exterior surfaces of the SiteStand.
Mix the disinfectant solution compatible with your system according to the label instructions
3
for the solution strength. Theracide disinfectant is recommended.
Wipe the SiteStand surfaces with the disinfecting solution, following disinfectant label
4
instructions for solution strengths and disinfectant contact duration. Ensure the solution
strength and duration of contact are appropriate for the intended clinical application.
Air dry or towel dry the SiteStand with a clean cloth according to the instructions on the
5
disinfectant label.

Cleaning and Disinfecting Peripheral Devices

For information about cleaning and disinfecting the peripheral devices, see the manufacturer's
operating instructions.
30
Chapter 4: Cleaning and Disinfecting Equipment
Tabla de contenido
loading

Tabla de contenido