R82 x:panda Guia Del Usuario página 29

Ocultar thumbs Ver también para x:panda:
Tabla de contenido
Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to
maintain a device in accordance with the instructions may invalidate the device's warranty. Furthermore,
failure to maintain a device may compromise the clinical condition or safety of users and/or their
caregivers. Do not perform service and maintenance while the user is in the product.
1st Day
• Read the User Guide carefully before
use and store it in a safe place for future
reference
• Store the enclosed tools on/at the product
Daily
• Use a dry cloth to clean the product
• Visual inspection. Repair or replace any
damaged or worn-out parts
• Ensure that all fasteners are in place and
fitted correctly
• Check all fittings and buckles for signs of
worn out parts
• Be aware of any MAX mark indications
before adjustment
Weekly
• Wipe off any residue and dirt from the
product using a cloth with warm water and a
mild detergent/soap and dry before use
• Disinfect the product. please check the
disinfection section in this User Guide for
more information
Monthly
• All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
• Oil the swing-away parts. We recommend
you use a professional lubricant system
Annually
• Inspect the frame for cracks or signs for
worn-out parts and carry out annual service.
Never use a product which appears to be
substandard or faulty
Washing
The cushion covers can be washed at 60°C
using a mild detergent in a washing machine. It
is not recommended to wash the foam.
Warewashing machine
This product can be washed at 60°C using a
mild detergent in 15-20 minutes in a washing
machine designed for medical devices. Use the
machine's drying function to dry the product.
Clean the accessories seperately.
Before washing remove the following items and
clean them separately:
• Accessories
• Cushions
support.R82.org
Disinfection
The product without accessories and cushion
can be disinfected with a 70% disinfectant IPA
solution. It is recommended to wipe off any
residue and dirt from the product, using a cloth
with warm water and a mild detergent/soap
without chlorine and let it dry before disinfection.
Service interval
A detailed inspection of the product must be
made every 12 months (every 6 months for heavy
users) and every time the product is reissued for
use. The inspection should be carried out by a
technically competent person who has been
trained in the use of the product.
The lifetime of this product in normal use is 5
years if all maintenance and servicing is carried
out in accordance with the manufacturer's
instructions and demonstrably recorded.
Service Check List:
Check and adjust the following functions as
per their intended use. Repair or replace any
damaged or worn-out parts. Please check the:
• Adjustment options for back, hip and legs
• Gas-spring for back recline and the security
lock
• Seat back is adjustable and can be locked
in position
• Nuts and bolts on the product are tightened
• Check all fittings and buckles for signs of
worn out parts
• Product for cracks or signs for worn-out
parts
• All labels on the product are intact
29
EN
Tabla de contenido
loading

Este manual también es adecuado para:

X:panda 1X:panda 2X:panda 3X:panda 4

Tabla de contenido