Troubleshooting List; Technical Service; Technical Data - Braun Aesculap ACCULAN GA 620 D Instrucciones De Manejo

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 37

10. Troubleshooting list

Fault
Motor runs too slowly or
has too low a torque
Motor does not run
No switching from right-
handed to left-handed
running

11. Technical Service

For service, maintenance or repairs contact your Aesculap/B. Braun
representative.
Carrying out modifications on medical equipment renders the
guarantee, and any licenses, null and void.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone: +49 7461 95-2700
Fax:
+49 7461 16-2887
Other service addresses can be obtained from the address indicated
above.
22
Cause
Ni-Cd battery is too weak
Ni-Cd battery is defective or flat
If the ACCULAN drill is used
infrequently and in addition the
applications are relatively brief, an
oxidation film may form on the
motor/battery connectors. This
oxidation film is an insulating layer
through which the low voltage may
not pass.
Gear or gear bearing defective
Switch faulty
Fault detection
No beep tone is heard from the
electronics when the trigger is
pressed
The beep tone from the
electronics can be heard when
the trigger is pressed
Slide on the battery cannot be
operated

12. Technical data

Machine type
DC voltage
Rated battery capacity
Rated current
Max. mechanical power
Max. torque
Feed
Dimensions (H x W x D)
Weight (with battery)
Type (according to IEC 601)
Spindle cannula
ACCULAN drill GA 620 D
Rectification
Charge battery
Check charger, if necessary replace
battery
Briefly turn the driven shaft by hand
(preferably with the adapter or
chuck mounted)
Repair by manufacturer
Repair by manufacturer
GA 620 D
9.6 V
0.8 Ah
3 A
100 W
3.6 Nm
0–900 rpm
225 mm x 140 mm x 55 mm
1150 g
B
3.4 mm
Tabla de contenido
loading

Tabla de contenido