To prevent wires from being pulled through the cable bundle, secure each
Caution:
wire close to where the outer black jacket stops prior to stripping cable insulation from
it.
8.
Remove approximately 75 mm (3") of cable jacket, shield, and filler strands.
9.
Cut off the excess cable core, making sure the cut is clean and square
10. Replace the steel spacer and tension block, then shorten the leads to approximately
40 mm (1.6").
11. Prepare the leads for soldering, and if heat shrink is used to insulate splices,
remember to slip on the heat shrink before soldering the leads.
12. Make the solder connections and position the heat shrink to completely insulate
each solder connection before shrinking it.
steel
spacer
tension
block
13. Remove any excess cable core, if necessary.
14. Apply PTFE type tape/sealant to all threads.
15. Add a pre-twist to the wires before screwing the probe sleeve and lower probe
assembly together: hold the probe sleeve still, and gently turn the lower probe
assembly counter-clockwise about 5 full turns. This avoids the wires being broken
when the probe and probe sleeve are assembled.
16. Screw the lower probe assembly clockwise into the probe sleeve, and tighten it with
a 17 mm (0.67") wrench.
17. Check that the instrument is operating correctly, using the test procedure on
page 25.
7ML19985JH04
Pointek CLS200/300 Standard – OPERATING INSTRUCTIONS