7. Troubleshooting

Trouble
Display direction
doesn't change
or changes
insensitively.
The SpO
2
and Pulse Rate
display unstable
Can not turn on
the device
No display
No display of the
wireless icon " "
Possible Reason
Maybe the oximeter
is not used for a long
time, the movable metal
ball within the
orientation-sensor
can not move freely.
1. The finger is not placed
far enough inside.
2. The finger is shaking
or the patient
is moving.
1. The batteries are
drained or almost
drained.
2. The batteries are not
inserted properly.
3. The device
is malfunctioning.
1. The device will power
off automatically when
it gets no signal for
8 seconds.
2. The batteries
are almost drained.
Hardware failure of
wireless transmission
function.
ENGLISH
Solution
Please shake the
oximeter with a certain
force to make the
movable metal ball
move freely.
If the problem still
exists, maybe
the orientation-sensor
is not working properly.
Please contact the
local service center.
1. Place the finger
correctly inside
and try again.
2. Let the patient
keep calm.
1. Change batteries.
2. Reinstall batteries.
3. Please contact the local
service center.
1. Normal.
2. Change batteries.
Please contact the local
service center.