Lovibond Tintometer MD 100 Manual De Instrucciones página 46

Ocultar thumbs Ver también para Tintometer MD 100:
Tabla de contenido
10 min
12
11
1
10
2
9
3
8
4
7
5
6
Wait for 10 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/l Phosphate appears on the display.
Implementation of the provision Polyphosphate with liquid
reagents
Select the method on the device
For testing of Polyphosphate HR with liquid reagents, carry out the described
digestion.
This test determines the content of inorganic total phosphate.The Polyphosphate
content arises from the difference between inorganic and ortho phosphate.
The test for total Phosphate LR with liquid reagents runs just as the test under Method
335, Phosphate HR with liquid reagents.
The result in mg/l anorganic Total Phosphate (ortho-Phosphate and Polyphosphate)
appears on the display.
Implementation of the provision total Phosphate with liquid
reagents
Select the method on the device
For testing of total Phosphate HR with liquid reagents, carry out the described
digestion.
This test determines all compounds of phosphorus present in the sample, including
ortho-phosphate, polyphosphate, and organic phosphorus compounds.
The test for total Phosphate HR with liquid reagent runs just as the test under Method
335, Phosphate HR with liquid reagent.
The result in mg/l total Phosphate appears on the display.
EN   Method Reference Book   1.0
Phosphate HR L / M335
EN
47
Tabla de contenido