Implementation of the provision Phosphate HR with fluid
reagent
Select the method on the device
14 ml
Filter approx. 14 ml sample
with a pre-rinsed filter (pore
size 0.45 µm).
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Hold cuvettes vertically
and add equal drops by
pressing slowly.
EN Method Reference Book 1.0
10 ml
24 mm
Fill 24 mm vial with 10 ml
prepared sample .
Zero
Press the ZERO button.
Add 25 drops
KS228 (Ammonium
Molybdate).
Phosphate HR L / M335
Close vial(s).
Remove the vial from the
sample chamber.
25
Close vial(s).