Accessories (available separately):
• ResMed Hospital Trolley power strip (ANZ, EUR, UK, USA)
• Cylinder mount
• Storage bin
• Circuit support arm.
For a full list of compatible devices, see the Ventilation Accessories
on www.resmed.com on the Products page under Ventilation
Devices. If you do not have internet access, please contact your
ResMed representative.
Assembly
Parts for fitting and assembly are shown in illustrations
A, B, C and E.
Fitting the IV support pole to the trolley
The Trolley arrives partly assembled; however the IV support pole
must be fitted.
1. Insert the IV support pole into the trolley table.
2. Secure in position using the M5 grub screw and 3 mm hex key.
Fitting the power strip
To power devices fitted to the Trolley, use the ResMed Hospital
Trolley power strip.
The power strip is a medical grade accessory validated for use with
the trolley. It improves trolley mobility and ease of use by providing
increased cable length and a single point of power connection/
disconnection for multiple devices connected to the trolley. The
power Sstrip is provided with double sided Velcro
CAUTION
• Ensure only Medical devices are connected to the power strip.
• Using a power strip other than the one provided by ResMed
may lead to a reduced level of safety.
• Use only power strips that comply with IEC60601-1 when
mounted in the trolley.
• Ensure the power strip is disconnected from mains power or
has enough cable length before moving the Trolley
To fit the Power strip:
1. Remove the Trolley rear panel.
2. Attach the power strip to the inner left side of the trolley centre
console (opposite the PSU Holders) using Velcro.
3. Thread the power strip cable through the lower cable port of the
trolley rear panel.
4. Replace and secure the trolley rear panel.
Using the Astral device with the trolley
CAUTION
Ensure the base plate is attached securely to the base of the
device and correctly fitted to the trolley table before use.
2
.
®