Technical specifications

Pressure–flow
curve
60
50
40
30
20
10
0
4
6
Dead space
information
Therapy
pressure
Resistance
Environmental
conditions
Gross
dimensions
Mask setting
options
The mask contains passive venting to protect
against rebreathing. As a result of manufacturing
variations, the vent flow rate may vary.
8
10
12
Mask pressure (cm H
Physical dead space is the empty volume of the
mask to the end of the swivel.
It is 93 mL.
4 to 20 cm H
O
2
Drop in pressure measured (nominal)
at 50 L/min: 0.2 cm H
at 100 L/min: 1.0 cm H
Operating temperature: +41°F to 104°F
(+5°C to +40°C)
Operating humidity: 15% to 95% non-condensing.
Storage and transport: -4°F to 140°F
(-20°C to +60°C)
Storage and transport humidity: up to 95% non-
condensing.
Mask fully assembled with elbow assembly – no
headgear.
5.43" (H) x 3.38" (W) x 3.46 " (D)
138 mm (H) x 86 mm (W) x 88 mm (D)
For S9: Select 'NASAL '
For other devices: Select 'STANDARD',
otherwise select 'VISTA' as the mask option.
14
16
18
O)
2
O
2
O
2
7
Pressure
Flow
)
(cm H
O
(L/min)
2
4
19
8
28
12
34
16
40
20
20
45