7 Conditioning
44
Operating Instructions Midwest Stylus
Please observe the manufacturer's instructions for using instrument
disinfectants.
7.2

Automated cleaning and disinfecting

NOTICE! Do not carry out automated cleaning of the quick couplings.
The cleaning and disinfection equipment used must be approved by its
manufacturer for the cleaning and disinfection of dental instruments and
comply with EN ISO 15883-1 (e.g. 95°C (203°F) and 10 min. holding
time).
For further details, refer to the operating instructions supplied with the
unit.
✔ The high-speed handpiece is conditioned with a cleaning and
disinfection device.
1. Check whether the high-speed handpiece is clean after conditioning
under good lighting (min. 500 lux) and color rendering index (min. 80
Ra).
®
®
Plus / Midwest
Phoenix ™ /Pro
6607290 D3673.201.01.03.02