5.6
Cleaning/disinfection
Product-specific safety instructions for the
reprocessing procedure
Damage to the product due to
inappropriate cleaning/disin-
fecting agents!
CAUTION
►
Damage to the product due to
inappropriate cleaning/disin-
fecting agents and/or exces-
CAUTION
sive temperatures!
►
►
►
►
Do not clean motors/handpieces by ultrasonic
treatment and do not immerse them in any liquids.
To avoid the risk of corrosion/malfunctioning,
allow any fluid that has entered the product to
drain out immediately.
Only use cleaning/disin-
fecting agents approved
for surface cleaning. Fol-
low the manufacturer's
instructions for the respec-
tive cleaning/disinfecting
agent.
Use cleaning and disinfect-
ing agents according to the
manufacturer's instruc-
tions which
– be approved for plastic
material and high-
grade steel,
– do not attack softeners
(e.g. in silicone).
Observe specifications
regarding concentration,
temperature and exposure
time.
Do not exceed the maxi-
mum permitted cleaning
temperature of 60 °C.
15