Validated procedure
Mechanical alkaline
cleaning and thermal
disinfecting
Special features
■
Use Eccos tray GB243R fitted
with Acculan 3Ti holders, or
mount appropriate holders in a
suitable tray (instructions for use
TA009721 for Aesculap Eccos
holder system).
■
Insert the product in its proper
position in the Eccos holder.
■
Connect the internal rinsing sys-
tem to the product/the Eccos
holder and the rinsing port of the
rinsing trolley.
■
If the washer/disinfector is not
equipped for internal rinsing, pre-
clean the cannulation manually,
Chapter Manual cleaning/disin-
fecting, phase I.
■
Otherwise, connect rinsing
adapter to the washer/disinfector
or rinsing cart.
■
Connect rinsing adapter to the
product: Screw rinsing adapter
into attachment 17.
Reference
Chapter Mechanical cleaning/disinfecting
and subsection:
■
Chapter Mechanical alkaline cleaning
and thermal disinfecting
17