Selection Of Measure Parameters; Measurement Screen; Storing Readings - Selecta STS-2011 L Manual De Instrucciones

Viscosimetros rotacionales
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

INSTRUCTION MANUAL CODE 80253
Configuracion
> Medidas
Memorias
Programacion
Opciones
Fig 7.2 Main menu
---Config Medida---
SP: L1
RPM:100.0
d: 1.0000
Max:
Fig 7.3 Measurement configuration
Fig 7.4 Measurement screen
------Midiendo------
SP: R1
RPM:
V:
425.4 cP
85 %
T: 031ºC
measurement % with
respects to the Eos
Configuration
Measurements
Memories
Programming
> Options
Fig 7.5 Main menu
------Options------
> Output
Info. manufacturer
Fig 7.6 Menu entry to save readings
J
.P. SELECTA s.a.
REV 0

7.3 Selection of measure parameters

Once the main options are selected, press «ON» and the measure screen will appear.
Use «TAB» to jump across the fields, and then «ENTER» to modify its value.
SP:
Indicates the spindle used in the measure.
RPM:
Indicates the speed used in the measure.
d:
Density.
Press «ENTER» key and
From the «SP» and «RPM» values, the «Max» field shows the EoS (End of scale).
Note : The viscosimeter only shows the spindles which are compatible with its model.
g/cm3
Note: Selectable speeds must into 18 pre-defined values.
60.0
IMPORTANT: Selecting a spindle which does not correspond to the one assembled will
cause measurement errors.
If we press the «QUIT» key, we'll return to the outlet configuration screen, losing all
data inserted in the measurement configuration.

7.4 Measurement screen

Data on screen is updated every reading is made (a reading is made every time the
spindle turns):
SP:
RPM:
V:
20
% :
The % factor indicates the reading percentage over the maximum measurable value
with the combination spindle - speed. This value should be the highest possible.
«QUIT»:
You can modify the rotation speed without leaving the measurement screen.
«START/STOP»
The following message appears:
Pause ends when pressing «START/STOP» again.

7.5 Storing readings.

El STS-2011 puede leer y almacenar lecturas de forma programada para posteriormente
descargarlas en un ordenador vía USB.
Inicialmente se programa la hora de inicio, fin e intervalo entre lecturas. Las lecturas
quedan almacenadas en el viscosimetro. Finalmente se descargan vía USB a un
ordenador.
Configurating data storing:
-
Check that main time its properly set. (See 6.3.1.3)
-
On main menu select «Opcions». (Fig 7.5)
-
Use keys «ENTER», «TAB» y
«Staus»:
«INI»:
«FIN»:
«INT»:
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: [email protected] - website: http://www.jpselecta.es
Febrero/2011
(It can be modified without notice)
Modify the density value only to meassure static viscosity.
For dynamic usual viscosity left at 1.0000 (Factory set)
to change its value.
Current spindle selected.
Revolutions per minute selected.
Viscosity measured.
In cP , mPa·s or in cSt (centiStocks) if the density has been entered.
EoS percentage.
EoS: End of Scale
Ends measurement.
Increases or reduces the spindle rotation speed (R.P.M.).
Pause during measurement. For instance, when trying to change
the sample.
«Motor stop»
to set :
ON / OFF to activate the data storing.
Initial time (HH:MM:SS) when storing readings starts.
End time (HH:MM:SS) when storing readings ends.
Scan time (HH:MM:SS) between readings.
Page: 18
Tel 34 937 700 877 Fax 34 937 702 362

Hide quick links:

Tabla de contenido
loading

Este manual también es adecuado para:

Sts-2011 rSts-2011 h100161110016121001613

Tabla de contenido