6.4 Changing mV to pH
pH-2003 allows measure on mV or pH.
On starting (On/Off), measure is in mV.
Pressing, «Mode» select «MEAS» to measure pH.
Measure in mV could be used to measure with an ion selective electrode.
6.5 Measure temperature
pH-2003 does not measures temperature. But it calculates a temperature compen-
sation according a temperature value entered from keypad.
In other words you can measure the sample temperature by a thermometer and
then enter it on the pH-2003
To enter temperature:
•
Select «ºC» pressing «Mode». (The units ºC blinks on the display)
•
Use keys «Up» or «Down» to select temperature.
•
Validate with «ENTER»
Note: If the temperature set it quite different fro the real temperature on sample,
the pH measured could have an error.
6.6 Calibrating pH-2003
El pH-2003 must be calibrated on 2 points.
Choose this two point by considering:
•
One point is always pH=6.98
•
If you are going to measure an acid solution choose pH=4.00
•
If you are going to measure an alkali solution choose pH=9.18
Calibrating pH-2003:
•
Select «STD1» press «Mode»
•
Inmerse the electrode cap on the pH=6.98 buffer.
•
Wait until the mV reading be stable.
•
Press «ENTER»
•
Clean and rinse the electrode cap with clean water.
•
Select «STD2» press «Mode»
•
Inmerse the electrode cap on the pH=4.00 or 9.18 buffer.
•
Wait until the mV reading se stabilize.
•
Press «ENTER»
•
Clean and rinse the electrode cap with clean water.
pH-meter is calibrated.
Note: Usually a pH-meter must be calibrated:
•
First time use
•
daily.
•
After change the electrode
•
After measuring strong acids (pH<2)
•
After measuring strong alkalis (pH>12)
•
After measure very dense solutions that can obstruct the membrane.