3. Special Care and Maintenance

Spindle cleaning:
Viscosimeters main maintenance is the spindle cleaning.
Please clean carefully all the sample rests from the spindle.
Handle the spindle with care in order not to bend them.
Keep the spindle and the motor shaft threads clean.
When changing the fluid, you must clean the spindle and its protector.
Motor shaft:
The viscosimeter motor shaft is a high precision piece, very sensitive to shock and
overexertion. That's the reason why it has a protective cap.
Assemble and dismantle the spindle carefully. The motor shaft is a very delicate
part.
Pleas protect the motor shaft when it's not in use.
Do not let the spindle assembled when it's not in use.
Spindle assembly:
When assembling the spindle to the motor shaft, block the shaft by pressing it
upwards before screwing the spindle. This prevents damages to the motor shaft.
The spindle is screwed left (the other way round a «normal» thread).
It's not necessary to «hold tight» the spindle to the motor shaft.
Do not use any tool to «tighten» the spindle to the motor shaft.
Do not use other viscosimeters spindles.
When changing the spindle, remember to select it in the viscosimeter.
Once the spindle is installed in the viscosimeter, do not turn it manually.
Once the spindle is assembled, do not lie the viscosimeter down.
General:
To turn the knob upwards and downwards, hold the viscosimeter with the other
hand.
Ambient temperature affects the viscosimeter measurement a lot.

4. Technical Specifications

Power supply:
Power consumption:
Unit dimensions: height/width/depth
Dimensions assembled:
Spindle number:
Error:
Repeatability:
«L» model measure range
«R» model measure range
«H» model measure range
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: [email protected] - website: http://www.jpselecta.es
Febrero/2011
(It can be modified without notice)
Page: 14
100/240 VAC 50/60Hz
22 VA
190 x 100 x 145 mm
620 x 360 x 230 mm
«L» model:
no 1, no 2, no 3, no 4
«R» model:
R2; R3; R4; R5; R6; R7
«H» model:
R2; R3; R4; R5; R6; R7
Spindle R1 is an accessory apart.
± 1%
0.2%
from 20
to
2.000.000 cP
from 100
to
13.000.000 cP
from 200
to
106.000.000 cP