KPA 01
Digital clamp multimeter
4. Connect test leads to circuit.
5. Current measurement will be shown on main display.
NCV Sensitivity and settings
1. Hold "FUNC" when you turn on the meter
and the following display will show up on the
LCD:
2. Push the "RANGE" button to adjust the NCV
sensitivity from 0-4. '4' is the highest sensitivity
where '0' will turn off the NCV functionality.
3. Push the "HOLD" button to save the selected sensitivity and return to
normal meter operation.
4. With the NCV function active (sensitivity between 1-4), if the sensor at the
top of the clamp gets close to a live conductor (AC voltage >90V) the NVC
indicator will flash and buzzer will sound.
Note:
1. Even though there is no indication, voltage may still exist. Do not rely
solely on NCV detector to determine the presence of voltage in a wire.
The measurement may be affected by the design of the outlet, type of
insulation and other external factors.
2. The voltage sensing indicator may turn on when a voltage is fed to the
meter.
3. Other external interference (ex. Flashlight, motor) may trigger the NCV
sensor.
Maintenance
Replacing the batteries
Warning
To avoid electrical shock, product damage or personal injury, remove test
leads before openning battery cover.