8- Maintenance & Calibration - VWR High Performance Guia De Inicio Rapido

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

VWR
HIGH PERFORMANCE PIPETTOR
®
Multichannel models
Volume
Model
range (µL)
8X-10µl
0.5-10
12X-10µl
8X-20µl
12X-20µl
8X-200µl
20-200
12X-200µl
8X-300µl
20-300
12X-300µl
The data given in the tables conform to the
ISO8655-2 Standard. With a precise pipetting
technique, the 2µL single channel model may be
used to aspirate volumes as low as 0.1µL and
the 10µL single channel model as low as 0.5 µL.
8 - MAINTENANCE & CALIBRATION
VWR recommends pipettor calibration and maintenance
at least once annually by the authorized service provider.
Please contact VWR directly.
12
Volume
(µL)
0.5
5
10
2
10
2-20
20
20
100
200
20
150
300
SPECIFICATIONS - Error limits
Systematic error
(µL)
(%)
± 0.09
± 17.6
± 0.22
± 4.4
± 0.22
± 2.2
± 0.11
± 5.5
± 0.22
± 2.2
± 0.40
± 2.0
± 0.55
± 2.8
± 1.10
± 1.1
± 2.20
± 1.1
± 1.10
± 5.5
± 1.65
± 1.1
± 3.30
± 1.1
Random error
(µL)
(%)
≤ 0.04
≤ 8.8
≤ 0.11
≤ 2.2
≤ 0.11
≤ 1.1
≤ 0.09
≤ 4.4
≤ 0.11
≤ 1.1
≤ 0.17
≤ 0.8
≤ 0.28
≤ 1.4
≤ 0.44
≤ 0.4
≤ 0.55
≤ 0.3
≤ 0.39
≤ 1.9
≤ 0.66
≤ 0.4
≤ 1.10
≤ 0.4
Tabla de contenido
loading

Tabla de contenido