Troubleshooting; Repair And Maintenance Of Vwr ® Rv 8; Cleaning; Spare Parts Order - VWR RV 8 Indicationes De Seguridad

Tabla de contenido

Troubleshooting

Any malfunctions during operation will be identified by an error message on the display.
Once a serious error message has been displayed, the lift moves to the top end position and the device can
no longer be operated.
Proceed as follows in such cases:
- switch off device using the main switch,
- carry out corrective measures,
- restart device.
Error code Cause
Temperature inside the
E03
device is too high
Target speed is not reached
E04
Motor blocked
Flash memory storage error Transmission error
E09
If the actions described fail to resolve the fault or another error code is displayed, then take one of the
following steps:
- contact your local VWR
- send the device for repair, including a short description of the fault.
Repair and maintenance of VWR
The instrument is maintenance-free. It is only subject to the natural wear and tear of components and their
statistical failure rate.

Cleaning

To clean the device use only water with a detergent that contains tensides, or use isopropylalcohol for
stubborn soiling.

Spare parts order

When ordering spare parts, please give:
- machine type,
- serial number, see type plate,
- software version,
- item and designation of the spare part, see

Repair

In case of repair the device has to be cleaned and free from any materials which may constitute a health
hazard. For this use the "certificate of compliance".
Effect
Room temperature > 40°C - Switch off the device
Load too large
Cable break
office for technical service,
®
®
Disconnect main plug prior to cleaning!
spare parts diagram and spare parts list.
Correction
- Let the device cool down
- Check the location
(see Product specifications)
- Reduce the volume in the evaporator
- Turn the unit off and on again
RV 8
28
Tabla de contenido
loading

Tabla de contenido