Installation; Maintenance & Servicing; Environmental Conditions - VWR 1000 Manual De Instrucciones

Agitadores de gran capacidad
Ocultar thumbs Ver también para 1000:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 20
I
nsTallaTIon
Upon receiving the VWR Large Capacity Stirrer, check to ensure that no damage
has occurred in shipment. It is important that any damage that occurred in
transport is detected at the time of unpacking. If you do find such damage the
carrier must be notified immediately.
After unpacking, it requires two (2) people to lift the Large Capacity Stirrer, from
the bottom of the unit, to place on a level bench, table or floor, away from explosive
vapors. Ensure that the surface on which the unit is placed will withstand typical
heat produced by the unit. Always place the unit on a sturdy work surface.
The Large Capacity Stirrer is supplied with a power cord that is inserted into the IEC
connector on the left side of the unit first, then it can be plugged into a properly
grounded outlet. The 120V unit plugs into a 120 volt, 50/60 Hz source. The 230V
unit plugs into a 230 volt, 50/60 Hz source.
M
& s
aInTenanCe
ervICIng
The Large Capacity Stirrer is built for long, trouble-free, dependable service.
No lubrication or other technical user maintenance is required. It needs no user
maintenance beyond keeping the surfaces clean. The unit should be given the
care normally required for any electrical appliance. Avoid wetting or unnecessary
exposure to fumes. Spills should be removed promptly. Do not use a cleaning
agent or solvent on the front panel or top plate which is abrasive or harmful to
plastics, nor one which is flammable. Always ensure the power is disconnected
from the unit prior to any cleaning. If the unit ever requires service, contact your
VWR representative.
e
C
nvIronMenTal
ondITIons
Operating Conditions: Indoor use only.
Temperature:
5 to 40°C (41 to 104°F)
Humidity:
20% to 80% relative humidity, non-condensing
Altitude:
0 to 6,562 ft (2000 M) above sea level
Non-Operating Storage:
Temperature:
-20 to 65°C (-4 to 149°F)
Humidity:
20% to 80% relative humidity, non-condensing
Installation Category II and Pollution Degree 2 in accordance with IEC 664.
EN
2
Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

2000

Tabla de contenido