Test number
Tests are automatically numbered by the megohmmeter to make their identification
easier. The test number is printed at the beginning of each test and it is stored in
memory.
Model and serial number
At the beginning of each test, the equipment model as well as its serial number are
registered, making it possible to relate the obtained results with their respective
equipment Calibration Certificate.
Tests definition
The MD-5060x megohmmeter model is an extremely versatile, which allows to
carry out several kinds of insulation tests automatically, recording all the results
in its internal memory or printing them. That's why it is necessary to define the
tests properly, selecting its parameters before starting the test:
•
Test voltage.
•
Equipment operation mode.
•
Minimum resistance limit, for the "Pass-fail test".
These three parameters shall be defined before starting the test. The test volt-
age is the only parameter that can be modified during the test.
The *,.0 keys makes it possible to enter the different equipment pro-
gramming modes. The 7 (decrease) and 8 (increase) keys select the value.
The functions of the *,.0 keys are mutually exclusive. Selecting one of
these keys implies releasing any other selected key.
The * and , keys select the test voltage programming mode in 500 V or
100 V steps. When one of these keys is on, the Decrease / Increase keys ad-
just the test voltage. Whenever the equipment is on, the ±500 voltage adjust-
ment function is on.
12