Technical Information

Inlet Screen Installation

Inlet Screen Installation
The inlet screen prevent the blades of the pump from being damaged
by debris greater than 5.2 mm diameter. The inlet screen does reduce
the pumping speed by about 10 %. The inlet screen is fitted in the
upper part of the pump, as shown in the figure.
Figure 43
The following figure shows the ISO 250 F pump flange section with
the protection screen fitted on it. As you can see, the overall
dimensions do not change as the inlet screen remains inside the
pump profile.
Figure 44
The system is shipped with the inlet screen mounted. Before using the system
NOTE
please check that the inlet screen in the correct position.
360/374
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01
(C)