Inlet Screen Installation - Agilent Technologies TwisTorr 804 FS Serie Manual De Istrucciones

Inlet screen installation

Figure 38
Figure 39
The inlet screens mod. 9699316 (ISO200-CFF10") and 9699350
(ISO250) prevent the blades of the pump from being damaged by
debris. The inlet screen, however, reduces the pumping speed as
follows:
9699316
9699350
TwisTorr 804 FS User Manual / 87-901-052-01
Reduction in pumping speed (%)
Ar
N
2
20%
18%
Technical Information

Inlet screen installation

He
7%
17
H
2
5%