When a data is indicated between two quotes ('...') it means that the indicated
NOTE
data is the correspond-ing ASCII character.
Tab. 5
Data Type
Logic (L)
Numeric (N)
Alphanumeric (A)
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01
<STX> (Start of transmission) = 0x02
<ADDR> (Unit address) = 0x80 (for RS 232)
<ADDR> (Unit address) = 0x80 + device number (0 to 31)
(for RS 485)
<WIN> (Window) = a string of 3 numeric char-acter indicating the
window number (from '000' to '999'); for the meaning of each
window see the relevant paragraph.
<COM> (Command) = 0x30 to read the window, 0x31 to write into
the window
<DATA> = an alphanumeric ASCII string with the data to be
written into the window. In case of a reading command this field
is not present. The field length is variable according to the data
type as per the following table:
Field Length
1
6
10
<ETX> (End of transmission) = 0x03
<CRC> = XOR of all characters subsequent to <STX> and
including the <ETX> terminator. The value is hexadecimal coded
and indicated by two ASCII character.
RS 232/RS 485 Communication Description
(C)
Technical Information
ValidCharacters
'0' = OFF
'1' = ON
'-', '.', '0' . . . '9' right
justified with '0'
from blank to '_' (ASCII)
323/374