Agilent Technologies Turbo-V 3K-G Manual De Instrucciones página 355

Tabla de contenido
23
1
°C
24
1
°C
25
1
°C
26
1
-
27
1
Bit Field
28
1
Sccm
Turbo-V 3K-G Pumping System User Manual / 87-901-011-01
0-100
Controller Power Section Temperature
Read the controller power section temperature.
If > 75 °C the controller goes in fail "controller over-
temperature"
0-100
Controller CPU Section Temperature
Read the controller CPU section temperature.
If > 80 °C the controller goes in fail "controller over-
temperature"
15-45
Pump Body Temperature Limit. Not used
0-6
B0-B3 – Status
0 = Stop
1= Waiting for interlock (interlock connections on remote I/O
connector missing)
2 = Ramp (the pump is increasing the speed)
3 = Auto tuning (the speed has been reduced because the
gas load is higher than the Power Limit)
4 = Braking (the pump is reducing the speed using the motor)
5 = Normal operation (the pump is at target speed)
6 = Fail (see error code byte Input Buffer Byte-27)
B4-B7 – Reserved
Error Code (bit field)
B7 – Too High Load (the gas load is too high for the pump)
B6 – Short Circuit (a short circuit happened between the two
of the three motor phases)
B5 – Over-voltage (not used)
B4 – Verify Line Select (the auto voltage selection mismatch
the real input voltage)
B3 – Power Fail (internal circuitry failure)
B2 – Controller Over-Temperature
("Controller Power Section Temperature" or "Controller CPU
Section Temperature" are over the limit)
B1 - Pump Over-temperature ("Pump Bearing Temperature"
or "Pump Body Temperature" are over the limit)
B0 – Check Connection To Pump
(the connection between controller and pump missing)
Flow Meter Reading
(Specific option, not available on standard models)
Technical Information
Profibus Message Mapping
(C)
17
222
211
164
205
206
223
355/374
Tabla de contenido
loading

Este manual también es adecuado para:

969-8888969-8889

Tabla de contenido