CONSERVATION AND CORRECTIVE AND PREVENTIVE
MAINTENANCE
Note: Use gloves and other protection systems during disinfection.
Spray lubrication
Fig.01.
Using the spanner wrench nut, remove
the Head of the counter-angle according
to (Fig.01).
- Press the spray oil keeping the bottle in the
upright position and apply until oil comes out on
the upper and lower parts of the counter-angle
head. (Fig.03).
- Cover the part with a paper towel to
remove the excess of oil.
- Handpieces not submitted to the
sterilization process shall be lubricated at least
twice a day.
For the perfect functioning of the
instrument, use only the lubricant
specified by Gnatus at the end of each
work period.
Lubricant
To proceed with the lubrication of your equipment, we recommend the use of the
product GTS Gnatus.
- The lubricant does not accompany the product.
- The non-utilization of the lubricant will void the product warranty.
The Ministry of Health registry for this product is made separately from the handpieces
since the product GTS is not manufactured by Gnatus.
Fig.02.
The lubrication shall be made through
the orifice using the adapter as illustrated
in (Fig.02).
Fig.03.
65
Adapter for
lubrication.