10 Care and control tasks for operators

E0804/ E0201
E0806
E0802
General error
General error
10 Care and control tasks for operators

General information

The
chuck should be removed and cleaned from time to time, and the dirt underneath the
handpiece tip removed with a brush.
Contamination on the plastic surfaces can be removed with a dry cloth.
Do not use aggressive cleaning agents!

Care and inspection plan

Care and inspection plan for QUBE II handpiece:
Interval
Weekly
Once a
month
26/32
No handpiece connected
Overload cut-off function
Handpiece blocked
Broken cable
The control unit is flashing red,
no display on the control satel-
lite, keys without function
Keys are not very responsive
motorhandpiece is designed for maximum durability. Despite this, the
Care and inspection tasks
Dismantle the tip and remove dirt
Dismantle the chuck, clean and lubricate
D21518
Original
Check whether the handpiece is
correctly connected - check plug
with bayonet connector!
Decrease contact pressure on
workpiece.
Check whether the chuck is closed.
Check smooth running of the
shaft.
Replace the handpiece cable.
Check connection cable from the
control satellite, remove the mains
plug and plug it back in again.
Unplug the mains plug and plug it
back in again.
Do not touch the control satellite
when restarting, so as to avoid
miscalibration!
Remarks
Do not use compressed air!
Lubricate with Art.No. 51/1
Instructions