Error
Vacuum release /
venting, too slow.
No vacuum being
generated.
Permanent vent-
ing noise.
Fluttering noise
from the silencer.
8.3 Error codes
The following indications will be given if
the electronic system detects an error:
•
A warning signal will sound
•
The display will flash alternately „Err" and
an error code.
End of error display:
•
Press control button (except for Err 3, 5
and 6)
•
Err 3, 5 and 6: Switch device off, follow
advice in section 2.6.2 "power failure".
With Err 5 it may no longer be pos-
sible to remove the mixing bowl.
With an error code as shown in the follow-
ing table, please proceed as advised.
Error code
Err. 2
Cause
•
Seal surfaces dirty.
•
Inlet filter plugged.
•
Silencer plugged.
•
Solenoid valve fault.
•
Venturi
nozzle plugged.
•
Suction filter system
plugged.
•
Silencer plugged.
•
Solenoid valve fault.
•
Bowl detection pin stick-
ing.
•
Solenoid valve fault.
•
Bowl detection pin stick-
ing.
•
Working pressure too
low.
Cause
•
Minimum vacuum level
not yet reached
•
Vacuum level has
dropped below 500
mbar.
•
Clean the seal surfaces (Sec. 3.2).
•
Change inlet filter (Sec. 3.5).
•
Replace the silencer (Sec. 3.6).
•
Connect filter pressure regulator in
series.
•
Have the device repaired.
•
Have the device repaired.
•
Have filter cleaned or changed
(Sec. 3.3).
•
Replace the silencer. (Sec. 3.6)
•
Connect filter pressure regulator in
series.
•
Have the device repaired.
•
Clean area around pin.
•
Have the device repaired.
•
Have the device repaired.
•
Clean area around pin.
•
Have the device repaired.
•
Check dynamic operating pressure
(see technical data, Sec. G.2).
•
Ventilate bowl and remove (see
section 2.6.1, points no. 2 and 3)
•
If the error recurs:
Have the device repaired.
- 9 -
Solution
Solution
EN