The fiber can be reset as normal, but the time of auto-fusion is too long.
Solutions:
Enter the menu of arc parameter; check whether the setup is efficient.
During arc test, the position of fiber's gap leans to one side of screen.
Solutions:
When the machine tests the arc, it will adjust the position of current and arc. If
the arc on the apparent of electrode were asymmetric, the arc position would
change. If the change was not too much, you can ignore it, otherwise please
mange it:
(1)
Enter the menu of maintenance, and then operate "Clean the
electrode" several times.
When you are merging or arc melting processes 1-5, which are configured in the
factory can not be used. They are too big or too small.
Solutions:
Because of electrode's aging, biggish change takes place in opposite position
between fiber and arc or operating environment.
1.- The condition of electrode's aging.
2.- Operate "Clean the electrode" if the top of electrode is well. Replace
electrode if the top of electrode is bad. (Consult 10<Maintenance>).
3.- Big change takes place in operation environment. Operations are as
follows:
(1)
Do arc test until "Moderate current" is displayed three to five
times.
(2)
Enter splice program the menu of arc parameter, and check the
value of arc current.
(3)
Translation motion decrease or increase current (premelting
current, splicing current).
(4)
Press
(5)
Press
02-2011
to save after confirming.
to exit menu.