1 Safety directions

Operating Manual 5201.6693.14 - 03.00
Before putting the product into operation make sure to read through and observe the
following safety directions as addition to the basic safety instructions.
Improper or negligent handling could result in death!
Since the standard-compliant configuration required by CISPR 16-1-2 and 16-2-1 and
also EN55016-1-1 and EN55016-2-2 means that it is not possible to limit the leakage
current to the value permitted by EN61010-1, and the base insulation required for a
category I protection device cannot be assured, it is imperative to provide additional
measures safeguarding against direct or indirect contact by the user.
The operator is responsible for ensuring that safety measures in accordance with
HD 384.4.41 S2 are applied during work with the line impedance stabilization network.
Before using for the first time, a secure connection must be made to the location's
grounding conductor. This connection must satisfy the requirements for a grounding
conductor connection in every detail. It must not be removed until after the network has
been disconnected from the mains power supply.
If the "fuse disconnector" protection method is used, a qualified electrical engineer
must indicate the points where grounding or a connection to the location's grounding
conductor may be sited.
The safety notes in the accompanying operating instructions and on the outside of the
device must be followed at all times.