Instruction: Set the permeate
totalizer watch
Step
1
2
3
4
Set block end time
Change the block end time so that the tank level has time to reach the volume set in the
tank level watch.
Step
1
2
3
UniFlux 10 Instrucciones de funcionamiento 28993587 AE
Action
In the Text Instructions tab, select the Base: SameAsMain instruction of the
Block: TMP_START_STOP block.
In the Instruction Box:
Set Breakpoint to 0.00 min.
•
In the Instructions: list, select Watch and Watch.
•
In the Parameters for field:
Set Signal to FIT142VTot.
•
Set Test to Greater_Than.
•
Set Value to volume of permeate.
•
Set Action to End block.
•
Click Insert.
Action
In the Text Instructions tab, select the End_Block instruction of the Block:
TMP_START_STOP block.
In the Instruction Box, set Breakpoint to a suitable time, for instance 60.00
min.
Click Change.
B Create a method with UNICORN 6 and above
B.2 An example of using the Method Editor
B.2.5 Phase 3: Fed Batch Concentration
253